Abstract: 
Purpose: To describe the case and the follow-up of an inadvertently damaged XEN gel implant during needling procedure.

Methods: Case report.

Results: A 60-year-old African American patient underwent an uncomplicated combined phacoemulsification with the insertion of a XEN gel implant. Two months postoperatively, a needling procedure was required. During the needling procedure, a fragment of the XEN gel implant was inadvertently damaged. One month postoperatively, the intraocular pressure was still controlled and the bleb was functioning well.

Conclusion: This case report reveals a previously unreported complication concerning XEN gel implant. It is important for the surgeon to pay attention during needling procedure, especially if there is a subconjunctival hemorrhage impairing the view, to defer the procedure until good visibility exists.
URI: https://scholarhub.balamand.edu.lb/handle/uob/5447
ISSN: 11206721
DOI: 10.1177/1120672118805876
Ezproxy URL: Link to full text
Type: Journal Article
Appears in Collections:Faculty of Medicine

Show full item record

SCOPUSTM   
Citations

12
checked on Nov 16, 2024

Record view(s)

54
checked on Nov 21, 2024

Google ScholarTM

Check

Altmetric

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.